Super Xenon Weather Meters
GX90 / GX75 / GX25
Light Source : Xenon Arc Lamp
Xenon Weather Meters are designed to reproduce outdoor and indoor deterioration factors such as sunlight, high and low temperature, rainfall and dew in a short period of time. The spectral power distribution of the Xenon Lamp gives the closest approximation of sunlight spectrum distribution, as well as long life with the best cost performance.
Related Standards : ISO 4892-1, ISO 4892-2, ISO 11431, ISO 12040, ISO 105-B02, ISO 105-B04, ISO 105-B06, ISO 2135, ISO 4665, ISO 3917, ISO/TR 18930, ISO 17398, IEC 60068-2-5, IEC 60068-2-9, IEC 61215, IEC 61646, ASTM D 6695, ASTM G 151, ASTM G 155, AATCC TM16, AATCC TM169, SAE J2527, SAE J2412, JIS K 7350-1, JIS K 7350-2, JIS K 5600-7-7, JIS A 1415, JIS A 1439, JIS K 5701-1, JIS L 0843, JIS L 0891, JIS H 8685-1, JIS K 6266, JIS D 0202, JIS D 0205, JIS E 4037, JEITA CP-3901, JIS Z 9107, JIS B 7754, JIS C 8917, JIS C 8938, JASO M 346, JASO M 351
*Subject to the accessories.
Xenon Weather Meter GX90
- World's No.1 Long-life Light Source 3000 hours.
*at 60W/m2(300-400nm), daylight filter, BPT63℃, 50%rh. - 9kW Water-cooled Xenon Arc Lamp of Long Life
- Specimen Capacity:126 pcs(150x70x1mm)
- Automatic Irradiance Calibration on the specimen surface as standard.
- Light Test: BPT63~89±2℃, Hum.50±5%rh(depends on the test condition)
- Dark Test:CAT38±2℃, Hum.95±5%rh
- Spray on both faces (Rain & Dew) is available.
- Wide variety of options and customization is available.
Xenon Weather Meter GX75
- 7.5kW Water-cooled Xenon Arc Lamp of Long Life
- Max. Irradiation: 120W/m2 at 300 to 400nm
- Specimen Capacity: 60 pcs(150x70x1mm)
- Light Test: BPT63~110±2℃, Hum.50±5%rh(depends on the test condition)
- Dark Test: CAT38±2℃, Hum.95±5%rh
- Spray on both faces (Rain & Dew) is available.
- Radiometer for Automatic Irradiance Calibration is included as standard.
- Saving energy by the temp. and humidity control system.
- Wide variety of options and customization is available.
Xenon Weather Meter GX25
- 2.5kW Water-cooled Xenon Arc Lamp of Long Life
- Compact model with excellent cost performance
- Max. Irradiation: 60W/m2 at 300 to 400nm
- Specimen Capacity: 96pcs(65x55x1mm)
- Light Test: BPT63~110±2℃, Hum.35~60±5%rh (at BPT63℃, 60Wm2)
- Dark Test: CAT38±2℃, Hum.95±5%rh
- Radiometer for irradiance calibration is included as standard.
- Wide variety of options and customization is available.
Standard Features
Name | Super Xenon Weather Meter | ||||
---|---|---|---|---|---|
Model | GX90 | GX75 | GX25 | ||
Light Source | 9kW Water- cooled Xenon Arc Lamp (WX9.0) |
7.5kW Water- cooled Xenon Arc Lamp (WX7.5) |
2.5kW Water- cooled Xenon Arc Lamp (WX2.5) |
||
Light Source Qty. | 1 | 1 | 1 | ||
Test Item | Light | 〇 | 〇 | 〇 | |
Light + Spray on front face (Rain) | 〇 | 〇 | Optional | ||
Dark | 〇 | 〇 | 〇 | ||
Dark + Spray on front face | 〇 | 〇 | - | ||
Dark + Spray on back face (Dew) | 〇 | 〇 | - | ||
Dark + Spray on both faces | 〇 | 〇 | - | ||
Irradiance | Wavelength 300 to 400nm *with daylight filter | 40~100W/m2 | 40~120W/m2 | 40~60W/m2 |
|
Temperature / Humidity Range |
Light | BPT *depends on the test condition |
63~89±2℃ | 63~110±2℃ | 63~110±2℃ |
Humidity *depends on the test condition |
50±5%rh | 50±5%rh | 35~60±5%rh | ||
Light + Spray on front face (Rain) | Chamber Temperature (CAT) | 38±2℃ | 38±2℃ | - | |
Dark | Chamber Temperature (CAT) | 38±2℃ | 38±2℃ | 38±2℃ | |
Humidity | 95±5%rh | 95±5%rh | 95±5%rh | ||
Dark + Spray on front face | Chamber Temperature (CAT)/Humidity | 40±2℃/95%rh or higher | 40±2℃/95%rh or higher | - | |
Dark + Spray on back face (Dew) | Chamber Temperature (CAT) | 38±2℃ | 38±2℃ | - | |
Dark + Spray on both faces | Chamber Temperature (CAT) | 38±2℃ | 38±2℃ | - | |
Specimen Capacity | 126 (150x70x1mm) | 60 (150x70x1mm) | 96 (65x55x1mm) | ||
Specimen Holder | Inclined type | 21 (per 4 stages) | 21 (per 3 stages) | 18 | |
Specimen Rack | Diameter | φ960mm (37.8in) | φ648mm (25.5in) | φ508mm (20.0in) | |
Rotation Speed | 2rpm | 2rpm | 3rpm | ||
Electrical Specifications | 50/60Hz | 3 phase, 400V, 65A | 3 phase, 200V, 55A | 3 phase, 200V, 43A | |
Earth Leakage Breaker (ELB) | ELB: 75A | ELB: 75A | ELB: 50A | ||
Water Consumptions | For wet bulb pot & humidifier | Pure water | TBA | approx. 5L/h | approx. 5L/h |
For specimen spray | Pure water (at 0.1MPa) | TBA | approx. 0.24L/min | - | |
For back spray | Pure water (at 0.1MPa) | TBA | approx. 0.24L/min | - | |
For lamp cooling water | Tap water or cooling tower water (at 25℃) |
TBA | approx. 6L/min | approx. 4L/min | |
For chamber temp. control / refrigerator |
Cooling tower water (at 25℃) | TBA | - | - | |
Outer Dimensions | Width x Depth x Height (including pipings) approx. | 148x175x205cm (58.3x68.9x80.7in) | 103x144x185cm (40.6x56.7x72.8in) | 100x129x180cm (39.4x50.8x70.9in) | |
Operating Weight | approx. | 1300kgs (2866lbs) | 540kgs (1191lbs) | 400kgs (882lbs) | |
Standard Accessories | Radiometer for Irradiance Calibration | 〇 | 〇 | 〇 | |
Irradiance Automatic Control System | 〇 | 〇 | 〇 | ||
Black Panel Temperature Automatic Control System | 〇 | 〇 | 〇 | ||
Refrigeration System for Chamber Temp. Control (water-cooled type) |
- | - | - | ||
Lamp Cooling Water Circulating System | 〇 | 〇 | 〇 |