Combined Cyclic Corrosion Test Chambers (Instruments)
CYP-90 / CYP-110 / CYP-120 / CYP-160 / CYP-200 / CYP-100R
Combined Cyclic Corrosion Test Chamber (Instrument) that can perform cycle tests that combine salt spray, drying, wetting, and outside air introduction, or these individual tests. It conforms to typical tests such as ISO, IEC, SAE, ASTM, JIS, JASO, and is widely used all over the world.
Combined Cyclic Corrosion Test Chambers (Instruments)
CYP-90 / CYP-110 / CYP-120 / CYP-160 / CYP-200
Test standard : ISO 9227, ISO 14993, ISO 16151, ISO 16701, ISO 21207, ISO 14993, ISO 11997-1, ISO 11997-2, ISO 19453-4, ASTM B368, ASTM B117, ASTM G85, IEC 60068-2-11, IEC 60068-2-52, JIS Z 2371, JIS H 8502, JIS G0594, JIS K 5600-7-1, JIS K 5600-7-9, JIS K 5621, JIS K 5981, JIS H 8681-2, JASO M 609, JASO M 610, JIS D 0201, JIS D 0202, JIS C 60068-2-11, JIS C 60068-2-52, and more.
Uniform distribution of the salt mist.
International standard ISO 9227 prescribes the salt mist rate 1.5±0.5ml/h at 80cm², while the CYP series enable better accuracy 1.5±0.2ml/h at 80cm².
Air Barrier Board is equipped to suppress PH fluctuations (PAT.).
The Air Barrier Board which is prescribed in the ISO 9227 is equipped in the solution supply tank to block outside air and suppress PH fluctuations.
Short time temperature shifting by the steam heating method (PAT.).
By heating inside the test chamber with a steam pipe, the temperature inside the chamber rises quickly which makes the temperature distribution inside the chamber uniform.
Automatically cleans the temperature sensor during the test (PAT.).
The resistance temperature detector installed in the test tank is automatically cleaned during the test to prevent salt and dirt from adhering and fluctuating the measured value during the test.
The solution recovery mechanism suppresses pH fluctuations in the solution (PAT.).
During tests other than the spray test, the solution is collected from the solution reservoir at the bottom of the spray tower into the solution replenishment tank to suppress pH fluctuations (CYP-120, CYP-160, CYP-200, CYP-100R).
Water Seal Method to seal completely
The lid and the chamber is sealed completely by water.
Slope of the ceiling / lid prevents water drops fall on the test specimen.
The lid and the chamber is sealed completely by water.
Easy-to-clean every corner
The rounded corners in the bottom of the test chamber make it easy to clean.
Transparent parts to check the operating status visually during the tests.
The transparent solution supply tank, air saturator, and the top lid allow you to visually check the remaining amount of solution or distilled water and the operating status of the chamber.
Combined Cyclic Corrosion Test Chambers (Instruments) CYP-100R
Rotation rack allows highest accuracy (PAT.).
A sample rotation mechanism that further pursues the accuracy of test results.
More uniform test results can be obtained than a testing chambers (instruments) in which a sample is fixed and placed in a test chamber. Pauses every 6 ° per min., (1 rotation every 1 hour), and the rotation direction reverses every rotation.
1 rotation every 1 hour (pauses every 6°/min) Reverse every 1 rotation
Standard Features
Combined Cyclic Corrosion Test Chambers (Instruments) | ||||||||
---|---|---|---|---|---|---|---|---|
Model | CYP-90 | CYP-110 | CYP-120 | CYP-160 | CYP-200 | CYP-100R | ||
Test item | Salt spray (neutral 5%) | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | |
CASS | - | - | - | - | - | - | ||
Dry | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | ||
Wet | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | ||
Immersion* | - | - | - | - | - | - | ||
Outside air introduction | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | ||
Low temp. | - | - | - | - | - | - | ||
Wet & High humidity | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | ||
Temperature & Humidity Range | Salt spray | Temp. | 35±1℃, 50±1℃ | 35±1℃, 50±1℃ | 35±1℃, 50±1℃ | 35±1℃, 50±1℃ | 35±1℃, 50±1℃ | 35±1℃, 50±1℃ |
Dry | Temp. | (RT+10)~70±1℃ | (RT+10)~70±1℃ | (RT+10)~70±1℃ | (RT+10)~70±1℃ | (RT+10)~70±1℃ | (RT+10)~70±1℃ | |
Humidity | 25±5%rh (at chamber temp.60℃) | 25±5%rh (at chamber temp.60℃) | 25±5%rh (at chamber temp.60℃) | 25±5%rh (at chamber temp.60℃) | 25±5%rh (at chamber temp.60℃) | 25±5%rh (at chamber temp.60℃) | ||
Wet | (RT+10)~50±1℃ | (RT+10)~50±1℃ | (RT+10)~60±1℃ | (RT+10)~60±1℃ | (RT+10)~60±1℃ | (RT+10)~70±1℃ | ||
Humidity | 60~95±5%rh (at chamber temp.50℃) | 60~95±5%rh (at chamber temp.50℃) | 60~95±5%rh (at chamber temp.50℃) | 60~95±5%rh (at chamber temp.50℃) | 60~95±5%rh (at chamber temp.50℃) | 60~95±5%rh (at chamber temp.50℃) | ||
Immersion | Temp. | - | - | - | - | - | - | |
Water level | - | - | - | - | - | - | ||
Outside air introduction | not controlled | not controlled | not controlled | not controlled | not controlled | not controlled | ||
Low temp. | Temp. | - | - | - | - | - | - | |
Wet & High humidity | Temp. | 50±1℃ | 50±1℃ | 50±1℃ | 50±1℃ | 50±1℃ | 50±1℃ | |
Humidity | 95%rh and above | 95%rh and above | 95%rh and above | 95%rh and above | 95%rh and above | 95%rh and above | ||
Specimen capacity (150x70x1mm each) | 58 pcs | 70 pcs | 104 pcs | 224 pcs | 272 pcs | 80 pcs | ||
Specimen mounting angle (against vertical) | 15°or 20° | 15°or 20° | 15°or 20° | 15°or 20° | 15°or 20° | 15°or 20° | ||
Chamber inner dimensions approx. (WDH) | 90x60x50cm (35.4x23.6x19.7 in) | 110x60x50cm (43.3x23.6x19.7 in) | 120x80x50cm (47.2x31.5x19.7 in) | 160x100x50cm (63.0x39.4x19.7 in) | 200x100x50cm (78.7x39.4x19.7 in) | 109x109x50cm (42.9x42.9x19.7 in) | ||
Capacity w/o lid approx. | 270L (9.5 cu ft) | 330L (11.7 cu ft) | 480L (17.0 cu ft) | 800L (28.3 cu ft) | 1000L (35.3 cu ft) | 594L (21.0 cu ft) | ||
Withstand load of specimen rack | 6 kgf (13.2lbf) | 7.5 kgf (16.5 lbf) | 11 kgf (24.3 lbf) | 20 kgf (44.1 lbf) | 25 kgf (55.1 lbf) | 11 kgf (24.3 lbf) (Rotation 1/4rph, Pause, Inversion) | ||
Electric capacity | 3 phase 200V, 16A (ELB: 20A) | 3 phase 200V, 16A (ELB: 20A) | 3 phase 200V, 37A (ELB: 50A) | 3 phase 200V, 42A (ELB: 50A) | 3 phase 200V, 42A (ELB: 50A) | 3 phase 200V, 39A (ELB: 50A) | ||
Tap water requirements approx. | 0.7m³/day | 0.7m³/day | 1.0m³/day | 1.2m³/day | 1.2m³/day | 1.2m³/day | ||
Pure water requirements approx. | 0.6L/h | 0.6L/h | 1.8L/h | 2.6L/h | 2.6L/h | 1.8L/h | ||
External dimensions approx.(WDH) | 183x101x148cm (72.0x39.8x58.3 in) | 203x101x148cm (78.0x39.8x58.3 in) | 200x108x190cm (78.7x42.5x74.8 in) | 270x128x190cm (106.3x50.4x74.8 in) | 310x128x190cm (122.0x50.4x74.8 in) | 158x166x189cm (62.2x65.4x74.4 in) | ||
External height with the lid open | 173cm (68.1 in) | 173cm (68.1 in) | 191cm (75.2 in) | 207cm (81.5 in) | 207cm (81.5 in) | 242cm (95.3 in) | ||
Operating weight approx. | 250kgs (551lbs) | 300kgs (661lbs) | 650kgs (1433lbs) | 750kgs (1653lbs) | 850kgs (1874lbs) | 550kgs (1213lbs) |
*Available immersion as an option : CYP-90D
**Available acid rain as an option : CYP-90A
***The specifications above are subject to change without notice.
Exhaust Treatment Unit (Optional)
EF-1R / EF-1RW / EF-2R
Completely remove the salt solution in the exhaust
Exhaust salt mist during the test is removed by passing it through the water spray. Treated water is circulated and reused by a pump, reducing consumption.
EF-1R for CYP-90, CYP-110, CYP-120
EF-1RW for CYP-160, CYP-200
EF-2R for large corrosion test chambers (instruments) CCT series.
Other Options for Corrosion Test Chambers (Instruments)
Exhaust Treatment Unit
Automatic Solution Supplying Unit
Automatic Pure Water Supply System to Air Saturators
Water Recycling Unit