Features

  • Capable of measurements in environments subject to high-frequency waves or high voltages.
  • A fiber optic thermometer offering a wide measurement range.
  • Analog output capability allows use for machine temperature control.
  • Measurement data can be imported into a PC using software* bundles with the thermometer.

* for Microsoft® Windows® 7, Windows® 8.1, Windows® 10 (for English)
* All PCs under the recommended environment are not guaranteed to work successfully with AMS-850.
* AMS-850 can be used only by the user that has a system administrative right (Adoministrator)
* AMS-850 does not support Mac OS.
* Microsoft® Windows® 7, Windows® 8.1, Windows® 10, and Excel® are registered trademarks of Microsoft Corporation in the United States and other countries.

Applications

Semiconductor / Electronics Industry

Semiconductor / Electronics Industry

Temperature measurements of PCBs in plasma for semiconductor manufacturing equipment or reliability testing of live electric circuit components

Power Industry

Power Industry

Accident prevention and maintenance of power transformer coils, high voltage lines, generators, and other power company facilities

Material Industry

Material Industry

Temperature control during heating, drying, disinfection of wood, ceramics, fibers, plastics, and other materials, or temperature measurement of microwave organic synthesis

Food Industry

Food Industry

Development of ready-to-ent microwave food or microwaveable packaging materials; temperature controls for disinfection or drying processes based on microwaves

Medical Industry

Medical Industry

Animal tests using hyperthermia equipment or MRI or NMR systems

Specifications

1ch. FL-2000 Specifications

Type FL-2000
Input point 1
Measurement range -195.0°C to 450°C
Resolution 0.1°C
Accuracy Calibration not executed -195.0 to 399.9°C : ±2°C, 400.0 to 450.0°C : ±5°C
Calibration executed ±0.5°C (within ±50°C of caribration point)
(at an ambient temp. of 25°C±5°C)
Temperature coefficient ±(0.01% of F.S./°C) within operating temperature
Operating environment Temperature : 0 to 40°C
Humidity : 15 to 80%RH(non-condensing)
Storage environment Temperature : -10 to 50°C
Humidity : 10 to 85%RH(non-condensing)
Display Height of character : 15mm RED LED
Display update 1s
Light source Blue LED
Power
supply
AC power Please specify a supply voltage when ordering
(choose from 100VAC, 115VAV, or 230VAC)
Dry battery six AA-size alkaline batteries
Power consumption approx. 12 VA
Memory UP to 15,000 data, one second interval
(synchronized with display)
Interface RS-232C Transfer rate : 9600bps
Data formation : 7bit, odd parity, 1 stop bit
Calibration One-point calibration
(using real temperature within a measured area)
Analog output Output rate 10mV/°C
Resolution 1mV
Accuracy ±(0.2% of F.S) at an ambient temperature of 25°C±5°C
Temperature coefficient ±(0.01% of F.S/°C)
External dimensions 210(W)×88(H)×230(D) mm
Protruded parts excluded
Weight aoorox. 3kg
Supplied accessory Power cable(two-wire plug attached)×1
Fuse(2A mini fuse)×1
Analog output cable×1
communication cable×1
PC software AMS-850 (CD ROM)×1
User's manual(main unit and software)×1

4ch. FL-2400 Specifications

Type FL-2400
Input point 4
Measurement range -195.0°C to 450°C
Resolution 0.1°C
Accuracy Calibration not executed -195.0 to 399.9°C : ±2°C, 400.0 to 450.0°C : ±5°C
Calibration executed ±0.5°C (within ±50°C of caribration point)
(at an ambient temp. of 25°C±5°C)
Temperature coefficient ±(0.01% of F.S./°C) within operating temperature
Operating environment Temperature : 0 to 40°C
Humidity : 15 to 80%RH(non-condensing)
Storage environment Temperature : -10 to 50°C
Humidity : 10 to 85%RH(non-condensing)
Display Height of character : 8mm RED LED
Display update 1s
Light source Blue LED
Power supply Please specify a supply voltage when ordering
(choose from 100VAC, 115VAV, or 230VAC)
Power consumption approx. 40 VA
Memory No memory
Interface RS-232C Transfer rate : 9600bps
Data formation : 7bit, odd parity, 1 stop bit
Calibration One-point calibration
(using real temperature within a measured area)
Analog output Output rate 10mV/°C
Resolution 1mV
Accuracy ±(0.2% of F.S) at an ambient temperature of 25°C±5°C
Temperature coefficient ±(0.01% of F.S/°C)
External dimensions 320(W)×99(H)×280(D) mm
Protruded parts excluded
Weight approx. 5kg
Supplied accessory Power cable(two-wire plug attached)×1
Fuse(2A mini fuse)×1
Analog output cable×4
communication cable×1
Conversion cable(RS-232C→USB)
PC software AMS-850 (CD ROM)×1
User's manual(main unit and software)×1

Fiber probes

How to order of Fiber probes

It can be chosen the length of cable and Teflon coating to suit the purposes of specific your application.
Replace the asterisk (*) in the model with selectable number.

order1 order2
number 1 Cable length
(Specify a cable length in 1m increments.)
2 2m
3 3m
... ...
number 2 Teflon coating length
(Specify a Teflon coating length in 0.1m increments, MAX.length : 1m)
01 0.1m
05 0.5m
10 1m

* The length of a fiber probe will be short after repar when the optical fiber is broken.

FS100-*M

Applications involving microwaves, high magnetic fields, certain food preparation processes, power lines, and electric circuit components.

FS100-*M

General-purpose type

temp. range : -190°C to 260°C

FS150-*M

The durable type which sensing part of FS100 made of the same thickness as the cable.

FS150-*M

High durability general-purpose type

temp. range : -190°C to 260°C

FS100H-**-*M

Oils or chemicals.

FS100H-**-*M

Liquid and chemical-resistant type

temp. range : -190°C to 260°C

FS150H-**-*M

The durable type which sensing part of FS100H made of the same thickness as the cable.

FS150H-**-*M

High durability liquid and
chemical-resistant type

temp. range : -190°C to 260°C

FS200-*M

Chemicals

fl_fs200.gif

Chemical-resistant type

temp. range : -30°C to 200°C

FS300-*M

Measurements performed in laser beam, plasma, or other intense light

fl_fs300.gif

Light-resistant type

temp. range : -190°C to 260°C

FS400-*M

Various materials, excluding liquids, in semiconductor manufacturing equipment or microwave heating devices

fl_fs400.gif

High temperature type

temp. range : 20°C to 400°C

FS500-*M

Semisolid substances, e.g., food

fl_fs500.gif

Needle type

temp. range : -60°C to 260°C

FS600-*M

Animal tests using hyperthermia equipment or MRI or NMR systems

fl_fs600.gif

Plastic type

temp. range : 0°C to 80°C

FS600 series is non-repairable.

Optional accessories

Extension cable FK1-*M

Extension cable FK1-*M

General-Purpose extension cable for normal use.

Extension cable FK2-*M

Extension cable FK2-*M

Heavy-duty extension cable for rough environments.

  • Replace the asterisk (∗) in the model number with a cable length in meters.
  • Extension cable is made of Silica fiber. The extension cable cannot be used with FS600-∗M(plastic fiber).
  • FA1(Optical adapter) is required for the connection of fiber sensors and relay cables.
  • The length of a fiber probe will be short after repar when the optical fiber is broken.

Optional adapter FA1

Optional adapter FA1

Con connect a probe to an extension cable.

Non-repairable

Compression fitting FN1-T / FN1-V

Compression fitting FN1-T / FN1-V

Compression fitting used to insert probe
into a vacuum device or pressure vessel.

FN1-T
(sealant material : TFE)
FN1-V
(sealant material : Viton)

Non-repairable

* Pressure rating : 1.33×10-4Pa to 11.0MPa at 20°C using stainless steel rod

Device for the microwave oven with a turntable

turntable

Special order product